Page 49 - Virtual Vascular Vol 6
P. 49
Lee Hoi Lam-OT.mp4
The aneurysm was opened without aortic clamping: The endograft was seen inside the aneu- The graft has some porosity issues and some small specks of blood was seeping out of the graft.
rysm sac. Most of the sac has thrombosed. This “Type IV” endoleak is a reason of continuous tension in the sac and sac expansion. The
other reason is endograft infection.