Page 45 - Virtual Vascular Vol 8
P. 45

IVC Filter: Failed Retrieval  “Double waveform” of jugular venous pressure














 IVC Filter retrieval.mp4
                   jvp.mp4

























































     Peak venous pressure occurs during atrial contraction “peak a”, and ventricular

     contraction “peak c” when the tricuspid value is closed and slightly bulging towards the
 While modern IVC Filters are often designed to be retrievable and thus temporary, retrieval
 can be met with difficulties.  Sometimes these filters will become permanent.  atrium.


 In this 57-year old patient a transfemoral retrieval of an IVC filter (Cordis Optease) place 2   Due to a change in intra-thoracic pressure, the effect of ventilation/ respiration in
 months ago was not successful.  Snared from below and withdrawn into a sheath, the tip   affecting the venous filling can also be seen.
 of the filter was adherent to the IVC wall.  Significant spasm of the IVC was noted and the
 retrieval was abandoned.