A 65 yr old male with acute renal failure. A large bore catheter was inserted via The position of the catheter is obviously too far to the left of the midline to be
the R transjugular route for haemodialysis. inside the SVC. Therefore inadvertent puncture of the common carotid artery
should be suspected. This is a serious injury due to large caliber catheter
CXR showed tip of catheter to left and not in superior vena cava as to be insertion, and proximity to the chest. A CT scan confirmed the iatrogenic injury.
expected. This was not realilzed until hemodialysis when arterial blood gushed
from the line.